Evaluation of the analytical performance of the Sysmex CS-2500 automated blood coagulation system

Breadcrumb

  • Home
  • Evaluation of The Analytical Performance of The Sysmex CS-2500 Automated Blood Coagulation System
  • Evaluation of the analytical performance of the Sysmex CS-2500 automated blood coagulation system

Touria El Bardi 1, 2, *, Moukaouim Mahassine 1, 2, Naïma Marzouki 1, 2, Salma Rouhi 1, 2, wafae Quiddi 1, 2 and Sanae Sayagh 1-2

1 Hematology Laboratory - Mohammed VI University Hospital, Marrakech, Morocco.
2 Faculty of Medicine and Pharmacy of Marrakech - Cadi Ayyad University, Marrakech, Morocco.
 
Research Article
GSC Advanced Research and Reviews, 2024, 21(02), 323–329.
Article DOI: 10.30574/gscarr.2024.21.2.0404
DOI url: https://doi.org/10.30574/gscarr.2024.21.2.0404
Received on 17 September 2024; revised on 11 November 2024; accepted on 13 November 2024
 
The aim of this study was to evaluate the analytical performance of the two new CS-2500 automated systems recently acquired by the laboratory. These are Sysmex automated systems that perform multiparametric haemostasis tests based on coagulometric, chromogenic and immunological principles with an optical detection system. The aim of this study was to evaluate their performance for prothrombin time (PT) and activated partial thromboplastin time (APTT) tests. Comparability results with the CS-2100 department's previous automated system showed high agreement between the systems, with high correlation coefficients (R ≥ 0.993). Intra-series and inter-series repeatability was less than 5%, demonstrating excellent measurement stability. The search for inter-sample contamination was negative. The automated systems showed excellent linearity over a wide range of values. All the results obtained during this evaluation point to very good analytical performance, making the CS-2500 suitable for medium- to high-activity haemostasis laboratories.
 
Automatic Coagulation Analyzer; Performance; Sysmex CS-2100 - CS-2500. 
 
https://gscarr.gsconlinepress.com/sites/default/files/fulltext_pdf/GSCARR-2024-…

Preview Article PDF

Touria El Bardi, Moukaouim Mahassine, Naïma Marzouki, Salma Rouhi, wafae Quiddi and Sanae Sayagh. Evaluation of the analytical performance of the Sysmex CS-2500 automated blood coagulation system. GSC Advanced Research and Reviews, 2024, 21(2), 323-329. Article DOI: https://doi.org/10.30574/gscarr.2024.21.2.0404

Copyright © Author(s). All rights reserved. This article is published under the terms of the Creative Commons Attribution 4.0 International License (CC BY 4.0), which permits use, sharing, adaptation, distribution, and reproduction in any medium or format, as long as appropriate credit is given to the original author(s) and source, a link to the license is provided, and any changes made are indicated.


All statements, opinions, and data contained in this publication are solely those of the individual author(s) and contributor(s). The journal, editors, reviewers, and publisher disclaim any responsibility or liability for the content, including accuracy, completeness, or any consequences arising from its use.